STRING COUNT=1 The Light Transfer File created by LTGEN and MOMGEN containing area statistics for calculating the system gain constant K and the noise floor.
STRING--OPTIONAL Mark-format data set with (line,sample) centers of rejected areas (see REJECT parameter).
STRING--OPTIONAL Output signal vs. noise plot file Print with PRINT/NOFEED option. E.g. DCL PRINT/NOFEED CCDNOISE.PLT
STRING--OPTIONAL Output tab-delimitted ASCII table file containing, for each exposure level: exposure time in msec signal at reticle 1, noise at reticle 1, computed at noise reticle 1, ...., signal at reticle 5, noise at reticle 5, computed at noise reticle 5 Reticle 1 is Upper left corner, 2 is upper right, 3 is lower left, 4 is lower right, and 5 is the center of the image.
REAL COUNT=0:2--OPTIONAL LIMIT=(loexp,hiexp) Only data in the Light Transfer File with exposures (in msec) between these values are used for the calculations.
INTEGER--OPTIONAL--For Galileo only Specifies exposure level number (1 thru number of inputs) at which extended exposure mode frames begins.
INTEGER--OPTIONAL
Reject areas with values differing from mean by more than 2-sigma.
REJECT=0 No area rejection
=1 Reject areas with bad system gain
=2 Reject areas with bad noise floor
=3 Reject areas with either term bad
KEYWORD--OPTIONAL. Specifies diagnostic printout.